Non-invasive photoacoustic computed tomography of rat heart anatomy and function

Non-invasive photoacoustic computed tomography of rat heart anatomy and function
a, Representative sketch of the 3D-PACT platform with a close-up view of the imaging aperture. b, Experimental setup for the rat heart imaging. The rat was mounted in a prone position, and the expanded light beam was directed towards the rat's chest from the bottom. c, Example of the signal synchronization between the 3D-PACT (top) and ECG (bottom) measurements. The 3D-PACT signal was collected from one transducer element on one array at multiple scanning positions (i.e., on the same latitude). In the photoacoustic signal diagram, the horizontal axis represents the scanning steps (i.e., time), and the vertical axis stands for the time-of-flight signal (data acquisition sampling at 20 MHz). The P, R, and T waves in the ECG signals correspond to atrial contraction, ventricular contraction, and ventricular relaxation, respectively. Credit: Li Lin, Xin Tong, Susana Cavallero, Yide Zhang, Shuai Na, Rui Cao, Tzung K. Hsiai, and Lihong V. Wang

Cardiovascular disease is the leading cause of morbidity, and preclinical investigations of animal models are critical for better understanding its mechanisms. Over the last several decades, non-invasive imaging of small animal models has provided in vivo insights into the structural and functional phenotypes with physiological and clinical relevance.

Photoacoustic computed tomography (PACT) is an emerging cardiac imaging technique that combines functional optical contrast from light illumination and from acoustic detection. However, the majority of the published PACT prototypes have not been demonstrated for non-invasive cardiac imaging of detailed anatomical or functional phenotypes due to suboptimal illumination and detection schemes.

In a new paper published in Light Science & Applications, a team of scientists, led by Professor Lihong V. Wang from the Caltech Optical Imaging Laboratory, Andrew and Peggy Cherng Department of Medical Engineering, Department of Electrical Engineering, California Institute of Technology, USA, and Professor Tzung K. Hsiai from the Department of Bioengineering and the Division of Cardiology, Department of Medicine, UCLA, and co-workers have presented the three-dimensional rat whole beating heart in vivo using a recently developed three-dimensional PACT (3-D-PACT) platform.

With regulated illumination and detection schemes, 3-D-PACT scanned the rat heart in 10 seconds and reconstructed a series of cardiac images based on a time-gating strategy. The 3-D-PACT significantly enhances the image clarity and captures the dynamic changes in the cardiac structure between obese and control rats; namely, chamber size, myocardial wall thickness, and intracardiac flow. Moreover, the 3-D-PACT is capable of recording the distinct blood flow in various sizes of the vascular system from healthy, obese, and hypertensive rats, including the aortas, pulmonary arteries, and left and right coronary arteries.

To address acoustic distortion from the chest ribs and lung air sacs that encompass the heart, the authors demonstrated a large acoustic detection aperture to collect photoacoustic signals from numerous view angles. Moreover, one of the most exciting (and challenging) goals in volumetric cardiac imaging is the heartbeat, which contains rich hemodynamic information but blurs the images. The authors applied time-gated reconstruction guided by the synchronized electrocardiogram (ECG) measurement to correct the motion artifacts. The 3-D-PACT, along with the ECG-aided synchronization, largely resolves the challenges from the previous PACT approaches and reveals the whole rat heart anatomy and hemodynamics non-invasively.

Non-invasive photoacoustic computed tomography of rat heart anatomy and function
a, Front view of the heart within a cardiac cycle. The heart is identified by a magenta circle at 4/11 T. BA, brachiocephalic artery; ITV, internal thoracic vessels; IV, intercostal vessels. b, Cross-sectional images of the heart on the sagittal plane. Each image is a maximum amplitude projection (MAP) of a slice marked by the green dashed lines in a (Time 0). c, Cross-sectional images of the heart on the coronal plane. Each image is an MAP of a slice marked by the yellow dashed lines in b (Time 0). LA, left atrium; LV, left ventricle; PuA, pulmonary artery; RA, right atrium; RV, right ventricle; SVC, superior vena cava. d, The same data (at Time 0) shown with color-encoded depths. Shallower structures were peeled away in the lower images to show the posterior anatomy. Credit: Li Lin, Xin Tong, Susana Cavallero, Yide Zhang, Shuai Na, Rui Cao, Tzung K. Hsiai, and Lihong V. Wang

The scientists summarize the advantages of their imaging setup: "To reveal the cardiac anatomy, 3-D-PACT employs a hemispherical acoustic detection aperture with large view angles to mitigate the influences induced by ribs and lungs; The 1064-nm light and uniform illumination also facilitate deep penetration of the entire heart; The periodic heartbeat allows the ECG-guided time-gating method to reduce motion-induced artifacts during a 10-second scan."

Successful imaging of the anatomical details further enables cardiac function and dynamics measurements. "The optical imaging contrast in 3-D-PACT depicts the distribution of myoglobin in the myocardium and hemoglobin in the blood, revealing cardiac anatomy and function with physiological relevance," they added.

"3-D-PACT is an open imaging platform for preclinical research and clinical translation by simultaneously providing cardiac anatomy and function. Compared with established imaging modalities, PACT of the heart offers additional physiology-related information with high imaging speed and resolution, yet without the need for ionizing radiation or invasiveness, making PACT a promising tool for cardiac imaging of human neonates," the scientists forecast.

More information: Li Lin et al, Non-invasive photoacoustic computed tomography of rat heart anatomy and function, Light: Science & Applications (2023). DOI: 10.1038/s41377-022-01053-7

Journal information: Light: Science & Applications
Citation: Non-invasive photoacoustic computed tomography of rat heart anatomy and function (2023, January 4) retrieved 12 May 2024 from https://medicalxpress.com/news/2023-01-non-invasive-photoacoustic-tomography-rat-heart.html
This document is subject to copyright. Apart from any fair dealing for the purpose of private study or research, no part may be reproduced without the written permission. The content is provided for information purposes only.